Hantek — 13.2 Impedance Settings

Utility EN signal is at or greater than the mark frequency, the sync signal is changed from TTL low level to TTL high level. ⚫ For N-cycle burst, when internal or manual are selected as the trigger source, the sync signal is TTL high level at the start of the burst. At the end of the specified count of cycles, the sync signal is TTL low level (it may not be zero intersection if the waveform has an associated starting phase). For an infinite-cycle burst, when manual triggered is selected, the sync signal is the same as that of the basic waveform. For all burst types, when external trigger is selected, there is no sync signal output.

13.2 Impedance Settings The Impedance setting affects the output amplitude and DC offset voltage. For the front panel [1] connector, the HDG3000C has a fixed series output impedance of 50Ω. If the actual load is different from the specified value, then the displayed voltage level will not match the voltage level of the components under test. To ensure the correct voltage level, ensure that the load impedance setting must match the actual load. Press [Utility] > CH1 Set > Imped softkey to choose HighZ or 50Ω . The default setting is HighZ. Impedance settings will be displayed on the screen. As shown in the figure below, the impedance of CH1 is set to "HighZ".

Figure 13.1 Impedance The generator will adjust the output amplitude and offset voltage automatically once the impedance setting is changed. For example, the current amplitude is 5Vpp and the output impedance is 50Ω. At this point, change the output impedance to HighZ and the amplitude displayed on the screen will double to 10Vpp. And the current amplitude is 5Vpp and the output impedance is HighZ. At this point, change the impedance to 50Ω, the amplitude will reduce to half of the previous value (2.5Vpp). Note: Only the displayed values change with the parameter and the real output Copyright © Qingdao Hantek Electronics Co., LTD HDG3000C User manual 73