Hantek — Chapter 5 Performance Index

Quick Guide Transient response time The output current from full load to half load, or from half load to full load, the time for the output voltage to recover to within 15mV is less than 50μs.

28 HDP44XX CH1 0.05%+10mV 0.2%+5mA 0.05%+10mV 0.2%+5mA CH2 CH3 0.1%+5mV 0.1%+10mA 0.1%+5mV 0.1%+10mA CH4 Mechanical Size 232*153*392mm Weight HDP43XX 9.15Kg HDP44XX 9.45Kg Power supply AC input (50Hz~60Hz) 100Vac±10%,115Vac±10%,230Vac±10%(Maximum 250Vac) Maximum input power 600VA Interface USB Device 1 piece USB Host 1 piece LAN 1 piece Digital IO 1 piece RS-232/485 1 piece (Option) GPIB 1 piece (Option) Surroundings Cooling method Air-cooled Operating temperature 0℃~50℃ Storage temperature -40℃~70℃ Humidity 0℃~30℃: ≤95% relative humidity 30℃~40℃: ≤75% relative humidity 40℃~50℃: ≤45% relative humidity Altitude Below 3000 meters