LIST:PACE? Function description Set or query the step rule between the channel lists. Parameter type: DWELL | TRIGGER. ch: Refer to OUTP command. Description DWELL: After executing the current item, automatically start the next item. TRIGGER: After executing the current item, wait for the trigger signal on the digital IO port to execute the next item. Return value Return DWELL | TRIGGER For example LIST:PACE TRIGGER,(@1) /* Set the channel 1 list step rule to TRIGGER */ LIST:PACE? (@1) /*Return TRIGGER */ Output at the End of the List : LIST:TERMinate:LAST Command format LIST:TERMinate:LAST LIST:TERMinate:LAST?
Function description Set or query the output of the channel after the execution of the list is completed. Parameter swi: OFF | ON | 0 | 1. ch: Refer to OUTP command. Description ON: When the list ends, the output voltage or current remains at the value of the last item in the list. OFF: When the list ends, the output voltage or current returns to the settings that were valid before the list was started. Return value Return OFF | ON For example LIST:TERMinate:LAST ON,(@1) LIST:TERMinate:LAST? (@1) List BOST Settings : LIST:TOUTput:BOSTep Command format LIST:TOUTput:BOSTep LIST:TOUTput:BOSTep? Function description Set or query the switch status of the BOST function of the list item.
Parameter 14