RUN key. IO: Digital IO port level trigger, that is, after pressing the RUN key, it will execute after waiting for the pulse list of the IO port. RMT: Remote command trigger, that is, after pressing the RUN key, wait for the LIST:TRIGger command to be executed. Return value Return KEY | IO | RMT For example LIST:TRIGger:SOURce KEY,(@1) /* Set the trigger source of the channel 1 list to key trigger */ LIST:TRIGger:SOURce? (@1) /*Return KEY*/ List Trigger Delay Time : LIST:TRIGger:DELay Command format LIST:TRIGger:DELay LIST:TRIGger:DELay? Function description Set or query the trigger delay time of the list. Parameter time: 0.0~3600.0s ch: Refer to OUTP command.
Description After the trigger signal occurs, the trigger delay timer starts timing, and the list starts to output after the timing ends. Return value Query return delay time, unit s. For example LIST:TRIGger:DELay 1.5,(@1) /* Set the trigger delay of the channel 1 list to 1.5s */ LIST:TRIGger:DELay? (@1) /*Return 1.5*/ Number of List Loops : LIST:REPEat:COUNt Command format LIST:REPEat:COUNt LIST:REPEat:COUNt? Function description Set or query the number of cycles of the list. Parameter nums: 0~9999, Integer. ch: Refer to OUTP command. Description If nums is set to 0, the list is output in a continuous loop, if nums is not 0, it represents the number of loops. 16