Test Sequence The test sequence has been designed to exercise the hardware one compo- nent at a time. Wherever possible, the system automatically detects proper component operation. Where the system cannot detect component opera- tion, the support engineer is required to confirm component operation (visu- ally or audibly). The test sequence is described below along with a brief description of each subtest. The sequence runs each subtest in order, unless an error is detected, in which case the system jumps to the reporting phase at the end of the test sequence, which displays the code for that error.
The ‘cycle length’ and ‘progress status’ LEDs are used to indicate progress through the test sequence. They should turn on one at a time, from top to bot- tom, left to right, one for each subtest. 27 Sub- test # Subtest Name Error Detection Subtest Summary Status Display 1 LEDs Visual Turn all LEDs ON, then OFF and repeat once more. Return LEDs to their initial state. This will display between every test (see below) All LEDs blink 2 Electronics self-test Automatic Perform at any power-up, not just diagnostic mode. The ‘warning’ LED is ON after power-up if this subtest ever fails. Topmost ’cycle length’ LED (long) ON and ’warning’ LED is ON.
3 Inlet Valve Audible Open and close valve multiple times. 2 ‘cycle length’ LEDs (long + medium) ON and ‘warning’ LED is ON. 4 Drain Valve Audible Open and close valve multiple times. All 3 ‘cycle length’ LEDs ON and ‘warn- ing’ LED is ON.