HP — Removing the cleaning tank assembly

Cleaning Tank Assembly Required Tools • #2 phillips screwdriver • Anti-seize compound • Tank chassis seal • Alcohol wipes Hardware • 8-32 x 3/8” phillips screw (x12) Removing the cleaning tank assembly 1. Remove the back panel. See “Removing the back panel” on page 39 . 2. Remove the user interface board. See “Removing the user inter- face board” on page 49 . 3. Using a #2 phillips screwdriver, remove the 4 tank holding clamps. Each clamp has 3 mounting screws securing each clamp to the upper housing. See Figure 39 . Figure 39: Cleaning tank clamp removal Tank holding clamps (X4) 4. Disconnect the 2 spade connectors from the drain solenoid. See Figure 40 . 5.