Figure 372: Measuring distances from outside edges of E and I 31. Measure and record the diameter of circle E along the center line between A and I. See Figure 392 Figure 373: Measuring diameter of circle E 32. Add the lengths derived from steps 30 and 31 (A-E and E-I) and then subtract the width of circle E derived from step 32. Record this total as “GLF” 33. Enter the Service Calibration menu by pressing the following button sequence on the display panel: 3,2,3,3,1 . See Figure 354. 34. Press the Edit Configuration button. 35. Press the GLeft button. Enter the value recorded from step 18 by press- ing the Increment , Decrement and Next Digit buttons. 36. When finished, press Done...
-369 NOTE: The value for GLeft must be between 5.250” and 5.750”.