HP — Performing part based calibration for HP Designjet Color 3D with the HP Designjet Diagnostics Software

Figure 391: Measuring distances from outside edges of E and I 32. Measure and record the diameter of circle E along the center line between A and I. See Figure 392 Figure 392: Measuring diameter of circle E 33. Add the lengths derived from steps 30 and 31 (A-E and E-I) and then subtract the width of circle E derived from step 32. Record this total as “Left Front” 34. Open HP Designjet 3D Diagnostic Software and select the Current Mod- eler from the pull down menu. 35. Open the Open Gantry Calibration Dialog box in HP Designjet 3D Diagnostic Software. 36. In the Part Calibration section, there are six boxes: LEFT, RIGHT, FRONT, REAR, LEFT FRONT and LEFT REAR.

Enter the measurements in the appropriate box. 37. Once the values have been entered, click on the green check mark. 38. Toggle the system power switch to accept the new values. 39. Run a test part to verify system is operating correctly. -379