Get/Send Calibration Files The .cal file contains the calibration values for the system. If the hard drive is replaced the .cal file will be automatically downloaded from the frame ID chip. Situations where the printer will not calibrate before a build, even if manual calibration has been performed by the technician, the factory .cal file may need to be obtained from Stratasys, Inc. Get Calibration The Get Calibration button allows a single calibration file to be retrieved from any printer. This calibration file will also be on the printer Frame ID chip. This file contains all the factory calibration information, including, but not limited to:
Tip Offset, Tip Depth, Y Lash, X Correction, and the table calibration .DAT file. When the button is pressed, a file selection dialog is presented allowing the assignment of a file name and the selection of a storage location. This file should be updated any time maintenance is performed that affects the factory calibration. Send Calibration The Send Calibration button allows you to send the complete set of factory calibrations to the printer as a single file. This includes but is not limited to: Tip Offset, Tip Depth, YLash, X Correction, and the gantry DAT file. This file is stored on the system frame ID chip. A calibration file can be retrieved from the modeler using Get Calibration .
When this button is pressed a dialog box appears that allows you to browse for and select the calibration file to send to the printer. XY Table Leveling Parts and Tools Required • Dial indicator • Head bracket • Basic tool set • Table leveling adjustment tool Checking the XY Table Level 1. Remove the substrate from the Z Stage table. 2. Mark a location on each corner of the platen surface with a marker, see Figure 401 These marks will be used to reference the same locations when leveling the table. -386