How to Perform an Offset Adjustment Calibra- tion 1 Obtain or create a calibration target (page 175) and enter the Calibration Menu in service mode (page 176). 2 Select the appropriate offset calibration you wish to perform (flatbed, ADF front or ADF back). 3 Load the target and follow the prompts on the front panel. Changes will be saved automatically when the adjustment is complete. How to Perform a Density Calibration Adjust- ment 1 Obtain a calibration target (part number PA03002-C261FJ) and enter the Calibration Menu in service mode (page 176). 2 Select Density Adjustment. 3 Load the target and follow the prompts on the front panel.
Changes will be saved automatically when the adjustment is complete. How to Perform an ADF Vertical Magnification Adjustment 1 Locate the page that shipped with the replacement ADF unit. 2 Enter the Calibration Menu in service mode (page 176). 3 Select ADF Vertical Magnification. 4 Enter the hexadecimal number from the page that shipped with the replacement ADF unit. C4265-90907 Adjusting the Copy Module 177