Downloaded from https://www.everyspec.com MIL-STD-1311D d. While simultaneously writing and erasing, adjust the unblanking level until there is no further increase in luminance of the area specified in the TSS. e. Without changing operating conditions, measure the dynamic equilibrium luminance at the center of the screen or at the location(s) specified in the TSS. 5. Precautions. a. To assure that stray light sources do not significantly affect luminance measurements, the light level recorded by the photometer in the measurement position shall be such that, with the TUT off, the photometer reading does not exceed 3 percent of the TUT luminance value to be measured. b.
Luminance measurements shall be made under carefully controlled conditions. Even under photometric laboratory conditions, accuracies within 5 percent are difficult to achieve. In the average production environment, accuracies within 10 percent may be expected. A major source of this error is the difference in ability of the different photometers to match the CIE (Commission Internationale de l'Eclairage) Standard Observer Luminosity Response. This error source is especially important when a large portion of the luminous energy is in the blue and red regions where deviations are most prevalent. 6. TSS requirements. a. The TSS shall specify the following: (1) Which method(s) are to be used.
(2) Operating conditions for the TUT. b. If values other than those specified in this test method, the TSS shall specify: (1) Photometer area (see 3a). (2) Deflection system timing (see 3b). (3) Erase system timing (see 3d). (4) Location(s) on screen if other than center, at which the particular measurement is to be made. 7. Comments on luminance testing. These comments are offered as an aid in meeting the requirement stated in 5a. a. Light shields. To obviate the necessity of a photometric laboratory, an "in situ" test setup can be accommodated if the ambient illumination can be sufficiently deleted from its effect on the photometric readings.
This desirable situation can best be obtained through the use of a light box as in figure 5412-1 or draped curtains. They may take the form of black painted cardboard in the shape of an inverted "U" channel or simply some black velvet cloth draped for best effect. The desired objective is to minimize the light leakage from affecting the photometer reading. It must be remembered that light leakage can be by direct illumination and by reflection. Small sources such as pilot lamps in the field of view of the photometer can introduce large errors. b. Intermediate baffles can be used to block direct light and also increase the leakage light distance.
This type of baffle would consist of a wall with a hole cut in it so as not to affect the desired signal light reading and placed approximately midway between the entrance pupil of the photometer and the test plane. The effect is that spurious light would have to travel a greater distance in order to register on the photometer. Figure 5412-2 shows how the light from the left extraneous source is blocked and also how the specular reflection of the right source is blocked. In addition any resulting diffuse reflection is greatly reduced. c. Photometer optics light scattering.
Light scattering in the photometer optics can introduce a large error when measuring a small dark area surrounded by a large bright area. A very small percentage of the light transmitted by the photometer optics is actually scattered rather than imaged as it should be in a "spot" photometer. If the surrounding luminance is much greater than the measured spot, the effect would be to bias the reading by means of the scattered light and the error could easily be of the same order as the true value. To reduce this effect, the use of target baffles can be employed. This may consist of an opaque material with a small hole cut into it.
The hole should be just larger than the area under test and placed as close to the test plane as convenient. This procedure will reduce the light scattering effect in proportion to the area reduction. METHOD 5412 5412-2 12 December 1972