Department of Defense MIL-STD-1311D — MIL-STD-1311D Test Methods for Electron Tubes — DoD Standard Manual PDF

Downloaded from https://www.everyspec.com MIL-STD-1311D METHOD 5434 AFTER IMAGE (Vidicons) 1. The tube shall be tested in a camera system with associated monitors as shown on the block diagram on figure 5426-1. 2. Align the tube in accordance with method 5437. 3. Using the spurious signal test pattern as shown on figure 5433-1, set up tube to specified conditions. Replace above test pattern by the "After image test pattern" as shown on figure 5434-1. The highlight brightness of the after image pattern shall be the same as the highlight brightness of the spurious signal test pattern. Line select the waveform monitor trace through the center information.

(Use IEEE response to reduce noise.) Adjust gain to make peak black-to-white signal level equal to 100 percent. Using the step gain switch, increase the signal amplitude to where a 3 percent signal level can be clearly seen. Expose the vidicon to the test pattern for an exact 30 seconds. Cap up the tube. Record the time (in seconds, using a stopwatch) required for the residual peak white signal to decay to a 3 percent level (above black) as after image. FIGURE 5434-1. After image test pattern. METHOD 5434 5434-1 28 March 1975