Downloaded from https://www.everyspec.com MIL-STD-1311D PLATEAU CHARACTERISTICS METHOD 6216 (Radiation counter tubes) 1. Plateau tracing (method A). When the plateau tracing (count rate versus applied tube voltage) of a radiation counter tube is made at the recommended operating voltage, at a counting rate of 100 Nps, and using a recording-rate meter having the following characteristics, the relative plateau slope, end-of-plateau voltage, and plateau length shall be within the limit(s) specified (see 1.1, 1.2, and 1.3, respectively). a. Resolving time. The resolving time shall be 50 ± 5 µ s. b. Discrimination level. The discrimination level shall be 1 ± 0.1 volts. c. Accuracy.

The recording-rate meter shall be capable of measuring uniform recurrence rates ranging from 10 to 200 Nps, with an accuracy of 5 percent. d. Deflection sensitivity. The deflection sensitivity of the recorder shall be not less than 1 inch per 50 Nps. e. Full-scale deflection. The full-scale deflection of the recorder shall indicate not less than 200 Nps. f. Voltage drive. The potential across the tube shall be applied at a uniform rate of increase not greater than 100 volts per minute. g. Time constant. The time constant shall be not greater than 2.5 seconds. 1.1 Relative plateau slope.

The relative plateau slope shall be determined from the average curve drawn through the plateau tracing obtained as specified in 1 above. A typical plateau tracing with an average curve drawn through it is shown on figure 6216-1. The relative plateau slope shall be determined over a voltage range specified. The relative plateau slope in percent per volt (see figure 6216-1) shall be given by 100 a/bc, where: a = Change in count rate (Nps) over specified voltage range. b = Specified voltage range (volts). c = Count rate (Nps) measured at the specified operating voltage. 1.2 End-of-plateau voltage.

The end-of-plateau voltage is the lowest applied counter tube voltage at which the relative slope of a line drawn tangent to the smoothed average curve (see figure 6216-1), obtained as specified in 1 above, exceeds a value given on the TSS. The end-of-plateau voltage is the lowest voltage at which the relative slope, 100 e/fg, is equal to the value given on the TSS, where: e/f = Slope of line drawn tangent to average curve (Nps). g = Absolute count rate measured at the point of tangency (Nps). 1.3 Plateau length. The plateau length of a counter tube shall be the difference between the end-of-plateau voltage and the plateau starting voltage. 2. Scaler (method B).

When plateau characteristics are to be determined by use of a scaler, details will be given on the TSS. METHOD 6216 6216-1 19 April 1968