Downloaded from https://www.everyspec.com MIL-STD-81OC 7.1.7.3 Determination of [Ib - Ia] and [Id - Ic] From figure 519,2-16 select at D = 48 inches; a ~f = .300 From figure 519.2-15 select at @f = .300 and f/f = a = fl,f = .233; ,,.. ,, 0 a value of I = .041 a 0 At f/~ = b = 2.0 select a value of Ib = 1.053 [Ib - ;a] = 1.012 substituting in eq. (11) At = 178.2 [1.012] = 180 3 gz From figure 519.2-17 select at D = 48 inches; a @f, = .51 From figure 519.2-24 select at 6f. = .511 and f/f = c = 0 value of I =D c 7; a At f/f = d = .222 select a value of Id = .435 0 Substituting in eq.

(12) As = 534.6 [.435] = 232.6 gz Add random noise A1+A:+A~= 17.82 g2 + 180.3 gz + 232.6 gz = 430.7 gz Add sines: 430.7 gz + 76.5 gz = 507.2 gz Take the square root: OARMS = (507.2)1’2= 22.5 grins 7.2 An A-7D utilizes a 20 mm (M61) cannon with a firing rate of 100 Hz. The instrument panel is shock mounted (4s Hz), weighs 75 pounds and the vector distance, D, is determined to be 49.6 inches from the nearest’mount of the panel. R~ is equal to 2 inches and the gun standoff distance is zero. Determine the gunfirevibration spectrum of an altimeter indicator rigidly mounted to the instrument panel. 519.2-21 METHOD 519.2