US Army Developmental Test Command TOP 1-2-621 — 9. DATA ACQUISITION, CALCULATION & REDUCTION.

Downloaded from https://www.everyspec.com TOP 1-2-621 6 February 2009 c. Prior to the characterization, before any sample is removed from any filter, filters will be weighed again inside their plastic bag. 7.2 Moisture Removal. a. A representative sample will be placed in a labeled beaker. The amount of sample used will depend on the total amount of sample collected and its physical properties. b. The samples will be placed in an oven and baked at 110 ˚C for two hours to remove any moisture. Desiccant material should be inside the oven to prevent the samples from absorbing moisture during the cool down period. c.

Let the samples cool down inside the oven and once at room temperature place it in a Ziploc plastic bag to avoid contact with moisture. If possible, store samples in a desiccator until the particle size distribution analysis can be performed. 8. PROCEDURE. a. Before every analysis, each sieve must be cleaned with a brush and/or compressed air and weighed. b. The sieves are nested one on top of the other in order of decreasing mesh size, with the largest mesh number on top and the collection pan on bottom. All sieves are used unless otherwise specified. c. Place the weighed sample in the top (first) sieve. d.

Activate the Meinzer II Sieve Shaker or equivalent device and shake the sieves for thirty minutes (see Figure B-2). NOTE: Samples containing very small particles (dust) have the tendency to remain in the sieves if the shaking process is too strong. Static electricity keeps the tiny particles attached to the walls of the sieves and to other bigger particles. To overcome this problem, it is helpful to use a micro-pulse shaker such as the Meinzer II Sieve Shaker. Be careful not to let small particles escape from the sieve setup during the shaking process. If necessary, the sieves can be sealed with adhesive tape.

Consider the weight added by the adhesive material to the sieves for calculation purposes. e. After the shaking process is complete, weigh each sieve to determine the weight of the retained sample. 9. DATA ACQUISITION, CALCULATION & REDUCTION. a. Record the weight of each clean sieve (without sample) before the test. B-4