5. Press and hold the volume reset soft key for at least 3 seconds to clear the error log. All the errors may not clear and that’s ok. 6. Press the soft key next to the icon to return to the “ User Screen 7. Turn on the suction by pressing the Vacuum System Power icon. 8. At this point, you can run all systems as if there were no errors. This will allow you to test Subsystems, individual components, check voltages, etc. 5-3 0000026959, Rev. C Effective Date: Dec 16, 2014 3:43:21 PM Print Date: Dec 16, 2014 04:03:57 PM