Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 3.5 Precautions. The low-pass filter is required in the circuit of directional coupler No. 1 to eliminate errors due to harmonic power passing through the directional coupler to the side arm with less than the correct attenuation. The filter should cut off above the highest frequency to be measured, but if possible below the second harmonic of the lowest frequency.

Since comparatively small differences of frequency and voltages are to be measured, the instruments used to measure these two should be capable of a resolution of 1 part in 10 4 , though the absolute accuracy of the voltage measurement need not normally be better than 1 percent. The voltage measurements are subject to error due to rectification of rf pickup, since the voltage of an electrode close to, or part of, the rf circuit must be measured. This effect tends to be more serious with high impedance instruments, which are usually the most precise. The errors can be reduced by the use of shielded leads and low-impedance circuits.

It may be necessary to use a digital voltmeter to obtain the required resolution, and to shunt the input with both capacity and resistance, to by-pass both rf and rectified currents. The shunting resistor will not affect the dc measurements as long as it is high compared to the lead resistance, and does not put a significant drain on the power supplied. This method may be used over any bandwidth, with appropriate choice of N, but is slower than method A in narrow-band cases. 3.6 Calibration. The system calibration is referred to the appropriate calibrations for power, frequency, and voltage, made by standard methods with the precautions indicated. METHOD 4277A 4277-5 25 March 1970