Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 3. Procedure. With the tube installed in the appropriate circuit, the line power is set at the specified level as observed on power detector No. 1. The arc loss, L a , shall be taken as the dB change in power at the termination, when the tube is replaced by a metallic short circuit. 4. Equipment. The 3 dB hybrid couplers used shall have a maximum terminated VSWR of 1.10 over the specified frequency range. Each coupler shall split the power evenly to within 0.25 dB and shall have a minimum isolation of 25 dB. 5. Formula. Although the loss may be read directly from the power meter, it may be also computed from: