Department of Defense MIL-STD-1311C — MIL-STD-1311C Test Methods for Electron Tubes – DoD Standard PDF

Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 METHOD 5246 FOCUSING (Cathode-ray tubes) The beam shall be focused at the center of the tube under the test conditions specified. Unless otherwise specified, for magnetic deflection types, deflecting yoke JEDEC No. 120 (see EIA-256-A), or equivalent, shall be used. 1. Focusing voltage at cutoff. With the control-grid voltage adjusted so that the pattern is barely visible, the voltage necessary to bring the beam to focus shall be within the limit(s) specified. 2. Focusing voltage at modulation condition. The voltage necessary to bring the beam to focus at modulation shall be within the limit(s) specified. 3.

Focusing ampere turns. JEDEC No. 106 focusing coil may be used. The number of ampere turns required for focus, as determined by multiplying the current in milliamperes by 4.5, shall be within the limit(s) specified. Alternately, JEDEC No. 122 focusing coil may be used. The number of ampere turns for focus, as determined by multiplying the current in milliamperes by 6.5 and by a correction factor 1.07, shall be within the limit(s) specified. The distance D from the center of the air gap of either focus to the reference line shall be as specified. METHOD 5246 19 April 1968 5246-1