Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 METHOD 5301 CURRENT (Phototubes) 1. Anode current. With the specified luminous flux incident on the photocathode, and with the specified anode voltages applied, the anode current shall be within the limit(s) specified. For gas phototubes, this measurement shall be made 30 seconds after the application of the anode voltages. 2. Dynamic anode current. A toothed wheel (calibrated with a stroboscope against a beat-frequency oscillator) shall be used to modulate a light beam from a specified source incident on the phototube cathode.
The speed of the toothed wheel shall be varied and the phototube output shall be fed into an amplifier which has been previously calibrated to insure its relative flatness of response over the desired frequency range. The dynamic sensitivity at each specified frequency shall be within the limit(s) specified. This test may be replaced by any suitable test demonstrated to be equivalent. 3. Dark current. Dark current is the current which flows between any two or more electrodes of a photoelectric device in the absence of such radiant energy as is normally used to excite the device during the test. The dark current, at the voltage specified, shall not exceed the limit(s) specified. 4.
Dynode current. With the specified potentials applied to the electrodes, the dynode current shall be within the limit(s) specified. METHOD 5301 5301-1 19 April 1968