Department of Defense MIL-STD-1311C — MIL-STD-1311C Test Methods for Electron Tubes – DoD Standard PDF

Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 4. Procedure. Luminance uniformity may be measured by use of any of the following four methods. The choice of a given method or methods is determined by the display application contemplated. The TSS should specify only the method(s) applicable to the end use. 4.1 Method A. This method measures the luminance uniformity factor (SW) under static writing conditions. a. Operating conditions shall be as specified in the TSS. b. Without simultaneously erasing, write the entire usable area specified in the TSS to saturation level. c. Stop writing.

Unless otherwise specified in the TSS, use a 1-second erase pulse and adjust its amplitude until application of a single pulse to the backing electrode will erase the last portion of a fully written area to just cutoff. d. Without simultaneously erasing, write the entire specified area until the brightest portion (found at the time of this measurement and within the specified area) is equal to the luminance level (Ba) specified in the TSS; then stop writing. e. Without changing the operating conditions, immediately measure the luminance (Bb) of the darkest area. f. Calculate the luminance uniformity factor (SW) for the specified value of (Ba) using the formula:

Luminance uniformity factor (SW) = Ba Bb Ba 4.2 Method B. This method measures the luminance uniformity factor (SE) under static erasing conditions. a. Operating conditions shall be as specified in the TSS. b. Without simultaneously erasing, write all of the area specified in the TSS to saturation level. c. Stop writing. Unless otherwise specified in the TSS, apply a 1-second erase pulse to the backing electrode; adjust its amplitude until application of a single pulse will erase the brightest remaining portion of the fully written specified area to the luminance level specified in the TSS (Bc).

If luminance level (Bc) is not achieved, readjust the erase pulse amplitude; repeat the writing step and erase. d. Stop erasing and without changing the operating conditions, find and immediately measure the lowest luminance level (Bd) within the specified area. e. Calculate the luminance uniformity factor (SE) for the specified value of (Bc) from the formula: Luminance uniformity factor (SE) = Bc Bd Bc 4.3 Method C. This method measures the luminance decay uniformity factor-dynamic decay write erase (DDWE) under conditions of dynamic writing and erasing in which the writing signal is generated by a random noise generator. The factor is a time ratio. a.

Operating conditions shall be as specified in the TSS. b. Without simultaneously erasing, write the entire area specified in the TSS to saturation level (noise not applied). c. Stop writing. Use the erase pulse width and frequency within the ranges specified in the TSS. Slowly increase the erase pulse amplitude until the last portion of the specified area has just been erased within time period (T2), as specified in the TSS. METHOD 5414 12 December 1972 5414-2