Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 METHOD 6211 STARTING VOLTAGE (Radiation counter tubes) The starting voltage is defined as the applied counter tube voltage at which uniform pulses of a specified average amplitude appear across the total resistance in the circuit shown on figure 6211-1. The test shall be performed using a calibrated high- impedance oscilloscope, a low-impedance power supply, and a high-impedance voltmeter (V). The count rate shall be less than 200 counts per second. The starting voltage shall be not greater than the limit specified. FIGURE 6211-1. Circuit used for measuring pulse amplitude and starting voltage.
METHOD 6211 19 April 1968 6211-1