Тексел ТХ01-RS — 9 Marking

3. By changing the DAC code and controlling the current (voltage) at the output of the analog device, to achieve the value of the output signal corresponding to the upper desired limit of the output current or voltage (in the general case, 20 mA or 10 V). 4. Commit the current DAC code as the upper desired current limit at the output of the analog device. 5. Set APPLY parameter to YES . 6. Exit the calibration menu. NOTICE To apply the calibration parameters, you need to edit both LdAC and HdAC . Calibration can be carried out using the service menu of the device or via the RS-485 network.

The commands used when calibration over the RS-485 network are given in Appendix A (section “Calibration”). Set the LdAC and HdAC parameters to achieve that the value of the output signal corresponds to a lower or upper desired limit of the output current (voltage). The first submission of these commands is considered the implementation of paragraphs 2 and 4 of the calibration algorithm, respectively. 9 Marking The housing of Device is marked with: • name of Device; • protection class of the housing according to GOST 14254; • voltage and frequency of supply; • power consumption; 52