Manual trigger: the instrument outputs pulse with over 1us pulse width from the beginning of pulse train. External trigger: the menu will hide the trigger output option which is enabled by external modulation connector (FSK Trig connector). This connector should not be used as external and internal at the same time. 3.3.9 Trigger Edge Trigger edge can be designated when external digital modulation interface (FSK Trig connector) is used as input. When it is used as input (i.e. internal trigger source), “rising edge” means that rising edge of external signal triggers output of a pulse train, and “falling edge” means that falling edge of external signal triggers output of a pulse train.
In gating mode, when polarity in parameter list is “positive”, external signal triggers output of a pulse train at high level, and “negative polarity” means that external signal triggers output of a pulse train at low level. When it is used as output (i.e. “internal” or “manual” trigger and trigger output is “ON”), the default edge is “rising edge”. If you need to change, press Param TrigEdge Fall (Gating mode:
press Param Polarity Negative ) Comprehensive Example First make the instrument run in pulse train mode, and then set a sine wave signal with period of 5ms and amplitude of 500mVpp as waveform of pulse train, set type of pulse train to be N-cycle, period of pulse train to be 15ms and recurring number to be 2.The specific steps are as follows: Use pulse train function Press Menu Burst Type NCyc successively (press soft key Type to select if Type is not highlighted) to set type of pulse train to be N-cycle mode. Figure 3 - 52 Set N cycle function Select waveform of Burst After setting N-cycle mode of pulse train, press Carrier Type Sine to select carrier signal as sine wave.
The default waveform of pulse train is sine wave, so it is unnecessary to change in this example. UTG2000B Series 105