UNI-T UTG2000B Series — 2.4.9 Set harmonic Wave

ratio: 1. Press Menu Wave Type Ramp Param 2. Select the parameter to enter edit mode, then input required numbers and select unit. Note: When enter symmetry degree value, press 50% label at the bottom of the display. 2.4.8 Set noise wave Default noise: Gauss noise with amplitude 100mVpp and DC offset 0mV. Steps for setting Quasi Gauss noise with 300mVpp amplitude and 1V DC offset: Press Menu Wave Type Noise Param to enter parameter editing mode. After setting, enter number and unit. 2.4.9 Set harmonic Wave Default frequency:1kHz, amplitude 100mVpp, DC offset 0mV, phase position 0°,harmonic wave type is odd order, total is 2 times, current is 2 times.

Amplitude 100mV, phase position 0°Steps for setting the harmonic wave type is all, wave time is 2 times, amplitude is 4Vpp, phase position is 0°: 16 UTG2000B Series