Press Clock source to select internal or external. Internal: Provide 10MHz clock source. External Receive external clock signal through 10MHz In (Requirement Frequency: 10MHz, amplitude is TTL electric level). If there is no detected valid external clock source, a prompt will pop up stating that “External 10MHz clock invalid” with symbol appears at the lower left corner. If the source is valid, will be shown instead. 3. Clock output Press Clock output to select ON/OFF When you turn on the clock output, 10MHz Out will output clock source (Frequency is 10MHz, amplitude is TTL electric level) for other devices Instrument Sync Method: 4.
Connect the 10MHz Out of the first instrument (clock output ON) to 10MHz In of the second instrument (clock source is external), set both instruments with the same output frequency. For more instrument synchronization, just connect the other instrument in cycle with the method above. 5. Sound Buzzer notification, press Sound and select ON/OFF. 6. Number separator Separator between numerical value, press Separator and select comma, space or null 7. Backlight Set the brightness of the display, press Backlight and select 10%, 30%, 50%, 70%, 90%. 8. Load Press Load to select factory setting or save setting 9. Save Press Save to save the current setting 10.
Arbitrary wave Save the arbitrary wave, and save to U disk. Press Arbitrary wave to select Internal or External (available on U disk insertion). Select Internal and you can see arbitrary wave, select External to enter into arbitrary wave input interface. 11. Software upgrade 24 UTG2000B Series