UNI-T UTG2000B Series — 3.1.3 Phase Modulation (PM)

1) Internal Source When modulation source is internal, modulation wave can be: sine wave, square wave, rising ramp wave, falling ramp wave, arbitrary wave and noise.After enabling FM function, the default of modulation wave is sine wave. If need to change, press Param ModWave in turn. Square wave: duty ratio is 50% Lead Ramp Wave: symmetry degree is 100% Tail Ramp Wave: symmetry degree is 0% Arbitrary Wave: Arbitrary wave length limit is 4kpts Noise: White Gauss noise 2) External Source When modulation source is external, carrier waveform will be modulated by an external waveform. PM phase deviation is controlled by ±5V signal level of external modulation input terminal on rear panel.

For example, if phase deviation value in parameter list has been set to 180˚, +5V of external modulation signal is equivalent to 180˚ phase shift. Lower external signal produce smaller deviation. Modulation wave Frequency Setting When modulation source is internal, frequency of modulation wave can be modulated. After enabling PM function, the default of modulation wave frequency is 100Hz. If need to change, press Param ModFreq in turn, and the modulation frequency range is 2mHz to 1MHz. When modulation source is external, carrier waveform will be modulated by an external waveform. The range of modulation signal input from external is 0Hz to 20Hz.

Phase Deviation Setting Phase deviation indicates the change between the phases of PM modulated waveform and the phase of carrier wave phase. Settable range of PM phase deviation is from 0˚ to 360˚, and the default value is 180˚. If need to change, press Param Phase Dev in turn. Comprehensive Example Firstly, make the instrument work in phase modulation (PM) mode, then set a sine wave with 200Hz from the internal of the instrument as a modulation signal and a square with frequency of 900Hz and amplitude of 100mVpp as a carrier wave signal. Finally, set the phase deviation to 200˚. Specific steps are seen as following: UTG2000B Series 41