UNI-T UTG2000B Series — 3.1.4 Amplitude Shift Keying (ASK)

Carrier Wave Frequency Setting See Carrier Wave Frequency Setting Modulation Source Selection The device can select internal modulation source or external modulation source. After enabling ASK function, the default of modulation source is internal. If need to change, press Param Source External in turn. 1) Internal Source When modulation source is internal, internal modulation wave is a square wave of 50% duty ratio (not adjustable). The ASK rate can be set to customize modulated waveform amplitude hopping frequency. 2) External Source When modulation source is external, carrier waveform will be modulated by an external waveform.

ASK output amplitude is determined by the logic level of modulation interface (FSK Trig connector). For example, output the carrier wave amplitude of current setting when external input logic is low, and output currently set carrier wave amplitude, when input logic is high, output amplitude is less than currently set carrier wave amplitude. ASK Rate Setting When modulation source is internal, frequency of ASK amplitude jump can be modulated. After enabling ASK function, ASK rate can be set and the settable range is 2mHz to 1MHz, the default rate is 100Hz. If need to change, press Param Rate in turn.

Comprehensive Example Make the instrument work in amplitude shift keying (ASK) mode, then set a logic signal with 300Hz from the internal of the instrument as a modulation signal and a sine wave with frequency of 15kHz and amplitude of 2Vpp as a carrier wave signal. Specific steps are seen as following: 35 3) Enable Amplitude Shift Keying (ASK) Function Press Menu Mod Type ASK in turn to start the ASK function. UTG2000B Series 47