UNI-T UTD2000 Series — UTD2000 Series User Manual

UTD2000 Series User Manual 1.3 Probe Compensation When connecting the probe with any input channel for the first time, it is required to make adjustment for this item, matching the probe with input channel. Probe without being compensated and corrected will lead to measurement error or mistake. In case of adjustment of probe compensation, follow the following steps: (1) Set the probe menu attenuation coefficient as 10Ч, place the switch on the probe at 10Ч and connect DSO probe with CH1. Connect the probe end and signal output connector to the probe compensator, the ground clamp and ground lead connector of probe compensator, turn on CH1 and press AUTO.

(2) Observe displayed waveform Over-compensation Correct compensation Under-compensation Figure 1-5 Probe Compensation and Correction (3) If screen displays “under-compensation” or “over-compensation” for waveform as shown in above figures, use screwdriver with nonmetal handle to adjust the variable capacitance on the probe until screen displays “correct compensation” for waveform as shown in the below figure. Warning: in order to avoid electric shock when measuring high voltage with probe, please ensure that the insulation lead of the probe is in good condition and do not contact the metal part of the probe when connecting high pressure power supply.

1.4 Automatic Settings of Waveform Display UTD2000 series DSO has automatic setting function. According to input signals, automatically adjust the vertical deflection factor, scanning time base and trigger mode until the most appropriate waveform is displayed. Apply automatic setting, the frequency of the measured signal is required to be 50Hz and the duty ratio is > 1%. Apply automatic setting: (1) Connect measured signal to signal input channel. (2) Press AUTO key. DSO will automatically set its vertical deflection factor, scanning time base and trigger mode.

If further careful observation is required, adjustment can be conducted again after automatic settings until the waveform display reaches required optimum effect. 1.5 Introduction to The Vertical System As shown in the figure below, there are a series of keys and knobs in the vertical 13