UTD2000 Series User Manual Chapter II Vertical System Each channel for UTD2000 series DSO can be set independently through the vertical system menu. After pressing CH1 or CH2 function key, the system shall display function menu of CH1 or CH2 channel. See Table 2-1 below for description. Table 2-1 Channel Menu 2.1 Setting coupling channel You will be learning through the following example to change the coupling of the input channel. If the measured signal connect to the CH1 channel is a sinusoidal signal containing DC component. If your press F1 to select as AC coupling, to set the CH1 into AC coupling mode. Then the DC component of the measured signal is blocked.
Waveform display is shown in figure below. Figure 2-1 Blocked DC Component of Signal If you press F1 to select as DC coupling, setting the CH1 to DC coupling mode. You should be able to see both the DC and AC components of measured signals at CH1 channel as shown in figure below. 16 Function Menu Setting Description Coupling AC Block the DC component of input signal. DC Pass through DC and AC components of input signal GND Disconnect input signal BW Limit ON Limit bandwidth to 20MHz to reduce displayed noise. OFF Turn off bandwidth limited Volts/Div Coarse Set vertical deflection factor by coarse adjustment based on 1-2-5 scale.
Fine Rough adjustment setting range is further subdivided for fine adjustments to improve vertical resolution. Probe 1× 10× 100× 1000× Please set the probe compensation system ratio to match the probe attenuation coefficient of the probe to be used. Invert ON Activate reverse function of waveform. Waveform will be displayed reversely OFF Waveform is displayed normally.