UTD2000 Series User Manual 11.3 Help Menu Press the HELP key to into help menu then press any key will display the help information of this key. 11.4 Upgrade program TheUSBupgrade program makes upgrade easier and more flexible. To use this function followthe steps below: (1) Check themodel name, software and hardware versions of your oscilloscopein Table 2-19. (2) Download the highest version of the update program document that matches your model, software and hardware versions online, and store it on the USB device (preferably a USB device recommended by us). (3) Power off the oscilloscope then insert the USB device. Then power on the oscilloscope again.

The prompt Please press F5 to continue or F1 to cancel appears on the screen as shown in the figure on theright Figure 11-2 USB update confirmation (4) When F5 is pressed, the prompt on the right will appear: “Are you sure you wish to update? If yes, press F5, otherwise press F1 to cancel upgrade”. (5) When the USB document is confirmed the following prompt will appear : “System upgrade is in progress. Please wait”. (This process may take 3 to 5 minutes) 46