UTD2000 Series User Manual Chapter XII Applicaion example Example 1: Measuring simple signals To observe and measure an unknown circuit signal, and to quickly display and measure the signal's frequency and peak-to-peak value. (1) To quickly display this signal, follow the steps below ① In the probe menu, set the attenuation factor to 10X and set the switch on the probe to 10X. ② Connect the CH1 probe to the circuitry point to be measured press AUTO. ③ The oscilloscope will carry out auto setup to optimise waveform display. In this status, you can further adjust the vertical and horizontal range until you get the desired waveform display.
(2) Voltage and time parameters for automatic measurement Your oscilloscope can automatically measure most display signals. To measure signal frequency and peak-to-peak value, follow the steps below : ① Press MEASURE to display the automatic measurement menu. ② Press F1 to enter the measurement type selection menu. ③ Press F3 to select Voltage ④ Press F5 to go to page 2/4, then press F3 to select measurement type : Peak value. Figure 12-1 Automatic measurement ⑤ Press F2 to enter the measurement type selection menu, then press F4 to select Time. ⑥ Press F2 to select measurement type : Frequency.
Peak-to-peak value and frequency measurement value are now displayed in positions F1 and F2 respectively. Example 2: Observing the delay caused by a sine wave signal passes through the circuit As in the previous scenario, set the probe attenuation factor of the probe and 48