Chapter 2 System Components Miniport , DEV=>::MAC->00-06-BE-00-09-4A::Baumer Optronic::VISIONSCAPE CMG20 192.168.254.2 00-06-BE-00-09-4A Microscan 0 0 VISIONSCAPE CMG20 After type="GigE_Camera" MACAddress="00-06-BE-00-09-4A"> TL=>GevTL , INT=>FD::MAC->00-1B-21-0A-0C- 72::National Instruments GigE Vision Adapter - Packet Scheduler Miniport , DEV=>::MAC->00-06-BE-00-09-4A::Baumer Optronic::VISIONSCAPE CMG20 192.168.254.2 00-06-BE-00-09-4A Microscan 0 0 VISIONSCAPE CMG20 We should now be able to bring up FrontRunner again and only see the one camera. We should also be able to run jobs from our original camera on this new device (assuming they are the same type of camera).
Everything looks fine in the debug window: 16:41:54 QuerySWRights: GigE license rights detected, features: Visionscape GigE (x8): YES / IntelliFind: YES / Third Party GigE Cameras: 0 allowed 16:41:56 CreateSystemsFromConfigFile: found 2 devices in dm.config file 16:41:56 Creating GigE System 1: GigEVision1 16:41:56 Channel 0 (Filter Driver): Microscan VISIONSCAPE CMG20 (1624x1236) MAC="00-06-BE-00-09-4A" IP="192.168.254.2" Connected 16:41:56 Creating SW System 1:
SoftSys1 16:41:57 Creating IO Server for System GigEVision1 16:41:58 Creating IO Server for System SoftSys1 This is certainly not as simple as it was when all you had to do to replace a camera was unplug the old one and plug in the replacement – but it’s not rocket science either. 2-48 Visionscape GigE Camera Guide