Hantek — HBT3000 series User manual EN20240309083423

Application measurement EN ⚫ When the number of valid data (excluding test anomalies) is 1, the sampling standard deviation and process capability index are not displayed. ⚫ When σn-1 is 0, Cp and CpK are 99.99. ⚫ The upper limit of Cp and CpK is 99.99. When Cp and CpK>99.99, it is displayed as 99.99. ⚫ When CpK is negative, CpK=0. Click on [STAT] -> Statistical Results -> Statistical List , and the statistical list will be displayed as follows: Figure 6-13 Statistical List Figure 6-14 Statistical List Compare the maximum and minimum values of voltage and resistance set in the comparator function with the statistical data.

If the statistical data is within the comparison range, the corresponding voltage or resistance will be displayed in white; If it is not within the comparison range, the corresponding voltage or resistance will be displayed in red; If the voltage and resistance values are both within the comparison range, the comparison result shows PASS; On the contrary, display UNPASS. The judgment results PASS and UNPASS (Hi, In, Lo of resistance and voltage respectively) can be output to EXT I/O. HBT3000 series user manual Copyright © Qingdao Hantek Electronic Co., LTD 34