Test and Diagnostic (P900 - P951) Note: These parameters are used for test and diagnostic purposes. P900 Software Revision Displays the EPROM (Flash ROM) software revision level. P901 Memory Tests the memory. Test is initiated by scrolling to the parameter or repeated by pressing ENTER. Display: PASS = normal FAIL = consult Siemens Milltronics. P907 Programmer Interface Tests the infrared communications link. Test is initiated by scrolling to the parameter or repeated by pressing ENTER. Parameters Display: PASS = normal FAIL = consult Siemens Milltronics. infrared communication link mirror BW100 P911 mA Output Value Displays the value from the previous measurement.
A test value can be entered and the displayed value is transmitted to the output. Upon returning to the RUN mode, the parameter assumes the actual mA output level. P940 Load Cell A, mV in Displays the mV signal input from the load cell. Range 0.00 - 60.00 mV. Page 62 Milltronics BW100 - INSTRUCTION MANUAL 7ML19985DJ01