Encoder diagnostics | Diagnosis for encoders with TTL/HTL/HTLs/ interface 8 Inverting the reference pulse You can invert the evaluation of the encoder's reference pulse on the product. Tap More The More menu is opened Tap Functions The Function settings screen is opened Tap the Invert the reference pulse option The evaluation of the reference pulse is inverted. Tap Back The changed settings take effect The Function settings screen is closed Evaluation options for switching signals Selection Function L1/L2 The rising and falling edges are evaluated for the L2/L/S2 switching signals. Select this setting if the encoder provides the L1 or L2 switching signals on separate signal pins.
H/L (Setting for encoder standard versions) Two rising edges are evaluated for the L2/L/S2 switching signals. This is indicated by the corresponding number- ing after the edge symbol. Select this setting if the encoder supports the Limit and Homing switching signals. H/L negated (Setting for special versions of encoders) Two falling edges are evaluated for the L2/L/S2 switch- ing signals. This is indicated by the corresponding number- ing after the edge symbol. Select this setting if the encoder supports the Limit and Homing switching signals. Setting evaluation options for switching signals Tap More The More menu is opened. Tap Functions The Function settings screen is opened.
Tap the desired mode of evaluation of the switching signals The switching signal evaluation was successfully selected. Tap Back The changed settings are applied. The Function settings screen is closed. 110 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024