heidenhain PWT 100, PWT 101 — 8.2 Performing Automatic Diagnosis

8 Encoder diagnostics | Performing Automatic Diagnosis 8.2 Performing Automatic Diagnosis If the Automatic Diagnosis menu is used, the product tries to automatically identify the encoder interface and run the appropriate diagnosis. The following procedure is used for the automatic diagnosis: Automatic Diagnosis No Encoder at X6? Encoder at X1? Yes Yes No Up = 24 V Up = 5 V Error message Manual Diagnosis Tap Automatic Diagnosis The Automatic Diagnosis is performed and, depending on the encoder interface, shows the PWT display , Online diagnostics or Level display screen The power supply at X6 is always set to U p = 24 V. The power supply at X1 is always set to U p = 5 V.

If you connect an encoder with a different supply voltage, you have to use the Manual diagnostics Further information: "Performing Manual Diagnosis", Page 60 If no encoder is connected to the product or if there are signal errors, the product displays an error message. In some cases, for example if the signal amplitudes are too low, the Automatic Diagnosis cannot clearly identify the interface of the connected encoder. In these cases, the Manual Diagnosis has to be called. The type and number of available screens vary depending on the connected encoder interface. Further information: "Various diagnostics screens", Page 58 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024 59