Encoder diagnostics | Diagnosis for encoders with serial interface 8 8.4 Diagnosis for encoders with serial interface 8.4.1 Online diagnostics screen 1 2 4 5 7 8 3 6 Figure 10: Online diagnostics screen 1 Status bars 2 Reset the minimum values 3 Referencing status 4 Position display 5 Mounting clearance 6 Position value 7 Revolutions 8 Status The Online diagnostics screen uses status bars to display the current diagnostic values of an encoder with serial interface. For the supported interfaces see "Information on the product", Page 10. The position values are displayed without interpreting any algebraic sign that may be applicable due to the interface definition.
On linear encoders, this may lead to the display of high position values. In this case the further analysis of the encoder should be performed with a PWM 21 phase meter and the ATS software. Alternately, you can evaluate the position values based on the position jump at the transition from zero to the maximum value. KCI support = PWT has calculated the value To obtain the value of the correct mating dimensions, a compensation calculation needs to be performed For more information, refer to the encoder documentation 62 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024