8 Encoder diagnostics | Diagnosis for encoders with serial interface 8.4.2 Temperature screen Figure 12: Temperature screen The Temperature screen displays the temperature values measured by the internal temperature sensor and the external temperature sensor of the connected encoder. Precondition: The encoder has an internal temperature sensor as well as an external temperature sensor and supports this function. If the connected encoder has only one temperature sensor or none at all, the product will display the Sensor is not supported message instead of a measured temperature value. If the temperature sensor is not connected or if there is a short circuit, a fault value will be displayed.
The fault value depends on the product and the corresponding interface. The display may indicate very high or very low values or a defined fault value of –300 °C, for example. Some encoders with EnDat 3 interface support several external sensors. Precondition: The encoder supports additional temperature sensors. The display will be adapted accordingly and a scroll bar is shown, if required. You can change the unit of the measured temperature values. You can select °C or °F . Further information: "General settings screen", Page 129 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2024 65