heidenhain PWT 100, PWT 101 — 8.5.7 Encoder status screen

8 Encoder diagnostics | Diagnostics for encoders with 1 V PP /11 µA PP /1 V PP + Z1 interface 8.5.7 Encoder status screen Figure 36: Encoder status screen (1 V PP /11 µA PP ) The Encoder status screen displays encoder errors. The Encoder status screen can be called from the following screens, using the Status button. PWT display screen Counts screen Switching signals screen Classification of encoder errors The messages displayed by the product for the 1 V PP and 11 µA PP interface are classified as follows:

Status screen Message Description Red Amplitude error Signal amplitude is too small ( PP or 3 µA PP ) or too large (> 1.35 V PP or 18 µA PP ) Red Counting error A counting error occurred (counting order Ua1, Ua2 is incorrect or the counting interval is too small) Red Invalid numeri- cal value between two reference marks An incorrect count value was detected while checking the position during reference-mark crossover. Example: When using a rotational encoder with a reference mark, a distance of 0 or the line count must always be detected between two reference marks. Red Reference mark not found A reference mark that was expected due to the current position value could not be identified.