Hantek — DPO7000 UserManual

To Trigger the Oscilloscope EN ⚫ Address: Triggered to search for the set address value, triggered on the read/write bit. After selecting the address as the trigger condition: ➢ Click the read/write direction key and select write, read, or read/write. When "8 Bits" is selected for the address bit width, this setting is not available. ➢ Click on the address bit width to select the desired address bit width. The address bit width can be selected from 7 bits, 8 bits, or 10 bits. ➢ Click on the address key and set the address value triggered by I2C through the pop-up numeric keypad. ⚫ Address and Data:

Triggered when the IIC address is received and the data (4 bytes of data) is equal to the user set value. (The oscilloscope simultaneously searches for the set address value and data value, and triggers when both address and data meet the conditions.) 5. Address Bits: Click on the Addr Bits width menu tab to select the desired address bit width, which can be selected as either 7Bits or 10Bits. 6. Address: Click on the Address menu label and set the address directly by rotating the multi-functional knob V0 or the pop-up numeric keyboard. 7. Data: Click on the Data menu tab and set Data 1, Data 2, Data 3, and Data 4 by rotating the multifunctional knob V0 or the pop-up numeric keypad.

DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 90