Hantek — DPO7000 UserManual

Measure EN 10.1 Square scale measurement Grid scale: This method can be used to quickly and intuitively estimate the frequency and voltage amplitude of the waveform, and simple measurements can be made through the division of the grid chart and the scale coefficient. For example, simple measurements can be made by calculating the relevant primary and secondary scale divisions and multiplying them by the proportional coefficient. If it is calculated that there are 6 main vertical scale divisions between the maximum and minimum values of the waveform, and the scaling coefficient is known to be 50mV/division, the peak to peak voltage can be calculated as follows:

6 divisions Ч 50mV/division=300mV 10.2 Measurement Parameter 10.2.1 Time Parameters 1. Period: defined as the time between the middle threshold points of two consecutive, like-polarity edges. 2. Frequency: defined as the reciprocal of period. 3. Rise Time: indicates the time for the signal amplitude to rise from the threshold lower limit to the threshold upper limit. 4. Fall Time: indicates the time for the signal amplitude to rise from the threshold upper limit to the threshold lower limit. 5. +Width: indicates the time between the threshold middle value of a rising edge to the threshold middle value of the next falling edge. 6. -Width: