System Utility Function Setting EN or negative pulse will be output from the auxiliary output connector. Set [Utility] -> Pass/Fail -> AUXOut to On, then the [Utility] -> More -> System -> AUXOut menu will automatically be set to "Pass/Fail". If the auxiliary output menu here is set to "trigout", then [Utility] -> Pass/Fail -> AUXOut function will automatically close. 13.6.3 Education model Click on the Education Model menu tab and select whether to turn Education Mode on or off. 13.6.4 Self-calibration The self-calibration program can optimize the oscilloscope signal path with maximum measurement accuracy.
You can run this program at any time, but if the ambient temperature changes by more than 5 ℃ , you should run this program. For more accurate calibration, please turn on the oscilloscope power and preheat for 30 minutes before performing self calibration. Follow the prompts on the screen to operate. To compensate for the signal path, disconnect any probes or cables connected to the input connector. Click on the SelfCal menu tab and select Offset Start . 13.6.5 Local upgrade [Utility] -> More -> System -> Local Upgrade , Click Local Upgrade and select whether to upgrade the firmware. If you click OK , select the firmware file to be upgraded; Click Cancel to exit the upgrade interface.
13.6.6 About this oscilloscope [Utility] -> More -> System -> About . Click on the menu label of About , and the system information of the oscilloscope will pop up on the screen, including the model, serial number, software version, hardware version, etc. of the oscilloscope. The menu also includes the functions included in the oscilloscope: 16 digital channels, any signal generator, digital voltmeter, power analysis, Bode diagram, histogram, RS232/422/485/UART, IIC/SPI, CAN LIN FlexRay, MIL-STD-1533 ARINC429, storage depth option. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 142