Decode EN 14.1 UART Decoding The RS232 serial bus consists of a transmit data line (TX) and a receive data line (RX). Figure 14.1 Schematic Diagram of UART Serial Bus The industry standard for RS232 uses "negative logic", which means that high levels represent logic "0" and low levels represent logic "1". Figure 14.2 Schematic Diagram of Negative Logic The baud rate is used in RS232 to represent the transmission rate of data (i.e. bits per second). RS232 requires setting the start bit, data bit, check bit (optional), and stop bit for each frame of data. ⚫ Start Bit: indicates when the data starts to be output. ⚫ Data Bit:
represents the actual number of data bits contained in each frame of data. ⚫ Check Bit: used to verify the correctness of data transmission. ⚫ Stop Bit: indicates when the data stops outputting. Click [Decode] in the button area to enter the decoding function menu. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 148