Troubleshooting EN 23 Troubleshooting The following is a list of possible faults and troubleshooting methods that may occur during the use of an oscilloscope. When you encounter these faults, please follow the corresponding steps to handle them. If you cannot handle them, please contact Hantek and provide the equipment information of your machine. 23.1 The display of waveform is ladder-like 1. The horizontal time base might be too low. Increase the horizontal time base to improve the display effects. 2. If the display type is "Vector", the lines between the sample points may cause ladder-like display results. 23.2 After connecting the power cord, the button light does not light up 1.
Check if the power switch of the oscilloscope is turned on. 2. Check if the fuse is intact. If a fuse needs to be replaced, please use a fuse that meets the specifications of this product. 3. If the button board still does not light up, please contact Hantek technical support department in a timely manner. 23.3 No waveform of the signal is displayed on the screen 1. Check if the probe is correctly connected to the BNC connector of the signal input channel. 2. Check if the channel is open (CH1~CH4 menu buttons). 3. Detect whether the signal channel to be tested has signal output. 4. If it is a DC signal and the amplitude is relatively large, please increase the amplitude range. 5.
You can press the automatic measurement button to automatically detect the signal first. If there is still no waveform display, please contact Hantai technical support department in a timely manner. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 235