Quick start EN Figure 4.4 To Use the Compensation Signal ⚫ Set the probe attenuation ratio to X10 and click on the button area [Auto Scale] . ⚫ Observe the pictures on the waveform to see if the square wave signal is displayed normally. As shown in the following figure. Figure 4.5 Square Waveform Signal ⚫ Check CH2~CH4 channels using the same method. If the square wave waveform displayed in the time base does not match the figure above, you can follow the instructions in the section on Probe Compensation . Warning: To avoid electric shock when using the probe, please make sure that the insulated wire of the probe is in good condition.