To Set the Vertical System EN 5.9 Amplitude Unit Open [CH1] , click on the Unit menu, and set the unit. The optional units are [W], [A], [V], [U], and the default unit is [V]. After modifying the amplitude unit, the units of functions related to the channel will also change. 5.10 Fine tuning Open [CH1] , click on the Fine adjustment menu, and the label will light up to indicate fine adjustment. The default setting is coarse adjustment. ⚫ Fine adjustment: The vertical gear will be adjusted within a small range, changing the vertical resolution to facilitate the observation of waveform details. ⚫ Coarse adjustment: Set the vertical gear in steps of 1 to 2 to 5.

5.11 Channel Delay When using an oscilloscope for actual measurements, the transmission delay of the probe cable may cause significant errors (zero offset). This series of oscilloscopes supports users to set a delay time to correct the zero offset of the corresponding channel. Zero offset is defined as the offset of the intersection point between the waveform and the trigger level line relative to the trigger position. Click on [CH1] , click on the Channel Delay menu, and a numeric keyboard will pop up to set the delay. The delay time can be set from -100ns to 100ns. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 39