To Set the Vertical System EN Figure 5.3 Zero Offset 5.12 Offset Cal When using an oscilloscope for actual measurement, the temperature drift characteristics of the device or external environmental interference cause a small deviation in the zero voltage of the channel, which affects the measurement results of vertical parameters. This series of oscilloscopes supports users to set a zero elimination voltage to correct the zero point of the corresponding channel, thereby improving the accuracy of measurement results. Click on [CH1] , click on the Offset Cal menu, and a numeric keyboard will pop up to set the numerical range.
5.13 Vertical Expansion This series of oscilloscopes supports two vertical expansion methods: center or GND, with the default being center: ⚫ Screen: When changing the vertical gear, the mathematical operation waveform will expand or compress around the center of the screen. ⚫ GND: When changing the vertical gear, the mathematical operation waveform will expand or compress around the zero position (grounding) of the waveform. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 40