To Set the Sample System EN 7.1 Timebase Mode Open [Acquire] , click on the Timebase Mode menu, and select time base modes: YT mode, XY mode, and ROLL mode. 7.1.1 YT Mode In this mode, the Y-axis represents the usual voltage and the X-axis represents the time. In YT mode, when the horizontal time base is set to 100 ms/div or slower, the oscilloscope enters scanning mode. In this mode, the oscilloscope first collects data on the left side of the trigger point, then waits for the trigger condition, and continues to complete the waveform on the right side of the trigger point after the trigger condition occurs. At the same time, the waveform data currently collected is displayed. Note:

Scanning mode observation signal frequency is low, it is recommended to set the "Channel Coupling" mode to "DC". 7.1.2 XY Mode This series of oscilloscopes supports the waveform display window "XY Window" in XY mode, where both the X-axis and Y-axis represent the voltage level. When the time base mode is selected as XY, the oscilloscope automatically opens channels 1 and 2, and channels 3 and 4 are automatically closed. Phase difference measurement: The Lissajous method can conveniently measure the phase difference between two signals of the same frequency. The following figure shows the measurement principle diagram of phase difference.

DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 48