To Set the Sample System EN Figure 7.1 Measurement Schematic Diagram of Phase Deviation According to or, where θ the definition of A, B, C, and D is the phase difference angle between channels, as shown in the figure above. Therefore, the difference angle can be obtained, which is: If the main axis of the ellipse is within the I and III quadrants, the obtained phase difference angle should be within the I and IV quadrants, that is, within (0 to π/2) or (3 π/2 to 2 π). If the main axis of the ellipse is within the II and IV quadrants, the obtained phase difference angle should be within the II and III quadrants, that is, within (π/2 to π) or (π to 3 π/2).
The X-Y function can be used to test the phase change generated by a signal passing through a circuit network. Connect the oscilloscope to the circuit and monitor the input and output signals of the circuit. Note: In XY mode, the oscilloscope will forcibly open CH1 and CH2 channels and close CH3 and CH4 channels. In general, longer sampling waveforms can achieve better display results, but due to storage depth limitations, longer waveform lengths mean a need to reduce the sampling rate. Therefore, in this measurement process, appropriately reducing the sampling rate can obtain a better display effect of the Lissajous pattern.
In XY mode, "Scan Mode", "Vector Display", "Protocol Decoding", "Acquisition Mode", "Pass/Fail Test", "Digital Channel", and "Afterglow Time" do not work. Using the Li Sha Yu method: 1. Connect a sine signal to CH1, and then connect a sine signal with the same frequency, amplitude, and phase difference of 90 ° to CH2. 2. Click on [AUTO Scale] , select the XY mode, and rotate the knob to adjust the sampling rate appropriately to obtain a better Lissajous pattern for better observation and measurement. 3. Adjust the vertical knobs corresponding to CH1 and CH2 channels to make the Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 49